Test ID C5FX C5 Complement, Functional, Serum

Useful For

Diagnosis of C5 deficiency

 

Investigation of a patient with an undetectable total complement (CH50) level

Method Name

Automated Liposome Lysis Assay

Reporting Name

C5 Complement, Functional, S

Specimen Type

Serum Red

Collection Container/Tube: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions:

1. Immediately after drawing the specimen, place the tube on wet ice.

2. Spin down and separate serum from clot.

3. Immediately freeze specimen.

Additional Information: Fasting preferred.

Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time
Serum Red Frozen 14 days

Clinical Information

Complement proteins are components of the innate immune system. There are 3 pathways to complement activation: 1) the classic pathway, 2) the alternative (or properdin) pathway, and 3) the lectin activation (mannan-binding protein: MBP) pathway. The classic pathway of the complement system is composed of a series of proteins that are activated in response to the presence of immune complexes. The activation process results in the generation of peptides that are chemotactic for neutrophils and that bind to immune complexes and complement receptors. The end result of the complement activation cascade is the formation of the lytic membrane attack complex (MAC).

 

Patients with deficiencies of the late complement proteins (C5, C6, C7, C8, and C9) are unable to form the MAC and may have increased susceptibility to neisserial infections. More than 30 cases of C5 deficiency have been reported. Most of these patients have neisserial infections.

 

Complement levels can be detected by antigen assays that quantitate the amount of the protein (C5AG / C5 Complement, Antigen, Serum). For most of the complement proteins, a small number of cases have been described in which the protein is present but is nonfunctional. These rare cases require a functional assay to detect the deficiency.

Reference Values

29-53 U/mL

Cautions

The total complement (CH50) assay (COM / Complement, Total, Serum) should be used as a screen for suspected complement deficiencies before ordering individual complement component assays. A deficiency of an individual component of the complement cascade will result in an undetectable total complement level.

 

Absent (or low) C5 functional levels in the presence of normal C5 antigen levels should be replicated with a new serum specimen to confirm that C5 inactivation did not occur during shipping.

Day(s) Performed

Monday through Saturday; Continuous with a 3 p.m. cutoff

Report Available

Same day/1 day

Performing Laboratory

Mayo Medical Laboratories in Rochester

Test Classification

This test has been modified from the manufacturer’s instructions. Its performance characteristics were determined by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information

86161

NY State Approved

Yes