Sign in →

Test ID CDOU Cadmium Occupational Exposure, Random, Urine


Specimen Required


Only orderable as part of profile. For more information see:

-CDUOE / Cadmium Occupational Exposure, Random, Urine

-HMUOE / Heavy Metal Occupational Exposure, with Reflex, Random, Urine


Secondary ID

608892

Useful For

Detecting occupational exposure to cadmium using random urine specimens

Method Name

Only orderable as part of profile. For more information see:

-CDUOE / Cadmium Occupational Exposure, Random, Urine

-HMUOE / Heavy Metal Occupational Exposure, with Reflex, Random, Urine

 

Triple Quadrupole Inductively Coupled Plasma Mass Spectrometry (ICP-MS/MS)

Reporting Name

Cadmium Occupational Exposure

Specimen Type

Urine

Specimen Minimum Volume

1.5 mL

Specimen Stability Information

Specimen Type Temperature Time
Urine Refrigerated (preferred) 28 days
  Ambient  28 days
  Frozen  28 days

Clinical Information

The toxicity of cadmium resembles the other heavy metals (arsenic, mercury, and lead) in that it attacks the kidney; kidney dysfunction with proteinuria with slow onset (over a period of years) is the typical presentation. Measurable changes in proximal tubule function, such as decreased clearance of para-aminohippuric acid, also occur over a period of years and precede overt kidney failure.

 

Breathing the fumes of cadmium vapors leads to nasal epithelial deterioration and pulmonary congestion resembling chronic emphysema.

 

The most common source of cadmium exposure is tobacco smoke, which has been implicated as the primary source of the metal, leading to reproductive toxicity in both male and female patients.

 

Chronic exposure to cadmium causes accumulated kidney damage. The excretion of cadmium is proportional to creatinine except when kidney damage has occurred. Kidney damage due to cadmium exposure can be detected by increased cadmium excretion relative to creatinine.

 

OSHA mandated (Fed Reg 57:42,102-142,463, September 1992) that all monitoring of employees exposed to cadmium in the workplace should be done using the measurement of urine cadmium and creatinine, expressing the results of mcg of cadmium per gram of creatinine.

Reference Values

Only orderable as part of profile. For more information see:

-CDUOE / Cadmium Occupational Exposure, Random, Urine

-HMUOE / Heavy Metal Occupational Exposure, with Reflex, Random, Urine

Cautions

Collection of urine specimens through a catheter frequently results in elevated values because rubber contains trace amounts of cadmium that are extracted as urine passes through the catheter.

Day(s) Performed

Monday through Friday

Report Available

2 to 4 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

NY State Approved

Yes