Test ID COVID Severe Acute Respiratory Syndrome Coronavirus 2 (SARS-CoV-2) RNA Detection, PCR, Varies


Ordering Guidance


Due to the non-specific clinical presentation of COVID-19 during the early stages of illness, testing for other respiratory tract infections (eg, influenza) may be warranted.

 

For the most up to date COVID-19 epidemiology and testing recommendations, visit www.cdc.gov/coronavirus/2019-ncov/index.html



Shipping Instructions


Ship specimens refrigerated (if less than 72 hours from collection to arrive at MCL) or frozen (if greater or equal to 72 hours from collection to arrive at MCL).



Specimen Required


Specimen Type: Nasopharyngeal (NP), oropharyngeal (OP ie, throat), nasal mid-turbinate, or nares/nasal swab

Supplies:

-Swab, Sterile Polyester, 10 per package (T507)

-Dacron-tipped swab with plastic shaft is acceptable

Container/Tube: Universal transport media, viral transport media, or equivalent (eg, Copan UTM-RT, BD VTM, MicroTest M4, M4-RT, M5)

Media should not contain guanidine thiocyanate (GTC).

For more information on acceptable transport media, see www.fda.gov/medical-devices/emergency-situations-medical-devices/faqs-diagnostic-testing-sars-cov-2

Specimen Volume: Entire specimen with a minimum of 1.5 mL (maximum 3 mL) of transport media.

Collection Instructions:

1. Collect specimen by swabbing back and forth over nasal or pharyngeal mucosa surface to maximize recovery of cells.

2. NP and OP swab specimens may be combined at collection into a single vial of transport media but only one swab is required for analysis.

3. Swab must be placed into transport medium. Swab shaft should be broken or cut so that there is no obstruction to the sample or pressure on the media container cap.

4. Do not send in glass tubes, vacutainer tubes, or tubes with push caps.

5. Do not overfill with more than 3 mL total volume of media.

 

Specimen Type: Nasopharyngeal aspirate or nasal washings

Container/Tube: Sterile container

Specimen Volume: Minimum of 1.5 mL

Additional Information: Do not aliquot into viral transport media, glass tubes, vacutainer tubes, or tubes with push caps.

 

Specimen Type: Lower respiratory tract

Sources: Bronchoalveolar lavage (BAL) fluid, bronchial washings, endotracheal aspirate, sputum

Container/Tube: Sterile container

Specimen Volume: Minimum of 1.5 mL

Additional Information: Do not aliquot into viral transport media, glass tubes, vacutainer tubes, or tubes with push caps.


Secondary ID

608825

Useful For

Detection of COVID-19 illness due to SARS-CoV-2

 

Recommended only for patients who meet current clinical and/or epidemiologic criteria defined by federal, state, or local public health directives: www.cdc.gov/coronavirus/2019-ncov/index.html

Method Name

Real-Time Reverse Transcription Polymerase Chain Reaction (RT-PCR)

Reporting Name

SARS Coronavirus 2 PCR Detect, V

Specimen Type

Varies

Specimen Minimum Volume

See Specimen Required

Specimen Stability Information

Specimen Type Temperature Time
Varies Frozen (preferred) 14 days
  Refrigerated  72 hours

Clinical Information

Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) is a positive-sense, single-stranded RNA virus that causes COVID-19. Like other coronaviruses that infect humans, SARS-CoV-2 can cause both upper and lower respiratory tract infection. Symptoms can range from mild (ie, the common cold) to severe (ie, pneumonia) in both healthy and immunocompromised patients. SARS-CoV-2 transmission occurs primarily via respiratory droplets. During the early stages of COVID-19, symptoms maybe nonspecific and resemble other common respiratory tract infections, such as influenza. If testing for other respiratory tract pathogens is negative, specific testing for SARS-CoV-2 may be warranted.

 

SARS-CoV-2 is likely to be at the highest concentrations in the nasopharynx during the first 3 to 5 days of symptomatic illness. As the disease progresses, the viral load tends to decrease in the upper respiratory tract, at which point lower respiratory tract specimens (eg, sputum, tracheal aspirate, bronchoalveolar fluid) would be more likely to have detectable SARS-CoV-2.

Reference Values

Undetected

Cautions

The US Food and Drug Administration has provided emergency use authorization of this test for testing human nasopharyngeal and oropharyngeal swab specimens.

 

The sensitivity of the assay is dependent on the timing of the specimen collection (in relation to symptom onset), quality, and type of the specimen submitted for testing.

 

The test is specific for SARS-CoV-2, and positive test results do not exclude the possibility of concurrent infection with other respiratory viruses.

 

Undetected (ie, negative) results do not rule out COVID-19 in patients and should not be used as the sole basis for treatment or other patient management decisions. Result should be correlated with the patient's history and clinical presentation.

Day(s) Performed

Monday through Sunday

Report Available

Same day/1 to 3 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test has received Emergency Use Authorization (EUA) by the U.S. Food and Drug Administration and is modified from the manufacturer's instructions with a bridging study. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information

87635

NY State Approved

Yes