Sign in →

Test ID F8INH Factor VIII Inhibitor Evaluation

Useful For

Detecting the presence and titer of a specific factor inhibitor directed against coagulation factor VIII 

Profile Information

Test ID Reporting Name Available Separately Always Performed
F8A Coag Factor VIII Activity Assay, P Yes Yes

Reflex Tests

Test ID Reporting Name Available Separately Always Performed
IBETH Bethesda Units No No
F8IS Coag Factor VIII Assay Inhib Scrn,P No No
CCCR Special Coagulation Interpretation No No

Testing Algorithm

Testing begins with coagulation factor VIII activity assay with dilutions to evaluate assay inhibition; if the factor VIII activity assay is decreased, an inhibitor screen will be performed to look for specific factor VIII inhibition. If specific inhibition is apparent, the titer of the inhibitor will be determined.

Special Instructions

Method Name

F8A: Activated Partial Thromboplastin Clot-Based Assay

F8IS, IBETH: Clot-Based Assay

Reporting Name

Factor 8 Inhib Prof

Specimen Type

Plasma Na Cit

See Coagulation Studies in Special Instructions.

 

Specimen Type: Platelet-poor plasma

Collection Container/Tube: Light-blue top (citrate)

Submission Container/Tube: Plastic vials

Specimen Volume: 3 mL in 3 plastic vials each containing 1 mL

Collection Instructions:

1. Specimen must be drawn prior to factor replacement therapy.

2. Spin down, remove plasma, and spin plasma again.

3. Freeze specimens immediately at ≤-40 degrees C, if possible.

4. Send specimens in the same shipping container.

Additional Information:

1. Double-centrifuged specimen is critical for accurate results as platelet contamination may cause spurious results.

2. If priority specimen, mark request form, give reason, and request a call-back.

3. Each coagulation assay requested should have its own vial.

Forms: If not ordering electronically, complete, print, and send a Coagulation Test Request Form (T753) with the specimen (http://www.mayomedicallaboratories.com/it-mmfiles/coagulation-test-request-form.pdf)

Specimen Minimum Volume

2 mL

Specimen Stability Information

Specimen Type Temperature Time
Plasma Na Cit Frozen 14 days

Clinical Information

Factor VIII inhibitors are IgG antibodies directed against coagulation FVIII that typically result in development of potentially life-threatening hemorrhage. These antibodies may develop in 1 of 4 different patient populations:

-Patients with congenital FVIII deficiency (hemophilia A) in response to therapeutic infusions of factor VIII concentrate

-Elderly nonhemophiliac patients (not previously factor VIII deficient)

-Women in postpartum period

-Patients with other autoimmune illnesses

Reference Values

FACTOR VIII ACTIVITY ASSAY

Adults: 55-200%

Normal, full-term newborn infants or healthy premature infants usually have normal or elevated factor VIII.*

*See Pediatric Hemostasis References in Coagulation Studies in Special Instructions.

 

FACTOR VIII INHIBITOR SCREEN

Negative

 

BETHESDA TITER

0 Units

Cautions

This test is not useful for detecting the presence of inhibitors directed against other clotting factors and will not detect the presence of lupus anticoagulants.

 

If presence or type of inhibitor is unknown, PROCT / Prolonged Clot Time Profile or LUPPR / Lupus Anticoagulant Profile should be ordered.

 

Occasionally, a potent lupus-like anticoagulant may cause false-positive results for a specific factor inhibitor (eg, factor VIII or IX). See preceding caution statement.

Day(s) Performed

Monday through Friday; Varies

Report Available

2 days

Performing Laboratory

Mayo Medical Laboratories in Rochester

CPT Code Information

85240-Factor VIII activity assay

85335-Bethesda titer (if appropriate)

85335-Factor VIII inhibitor screen (if appropriate)

85390-26-Special coagulation interpretation (if appropriate)

NY State Approved

Yes