Sign in →

Test ID FLEC Flecainide, Serum


Specimen Required


Collection Container/Tube: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 1.5 mL

Collection Instructions:

1. Draw blood immediately before next scheduled dose.

2. Centrifuge and aliquot serum into a plastic vial within 2 hours of collection.


Forms

If not ordering electronically, complete, print, and send a Therapeutics Test Request (T831) with the specimen.

Secondary ID

9243

Useful For

Optimizing dosage

 

Assessing flecainide toxicity

 

Monitoring compliance

Method Name

Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS)

Reporting Name

Flecainide, S

Specimen Type

Serum Red

Specimen Minimum Volume

0.5 mL

Specimen Stability Information

Specimen Type Temperature Time
Serum Red Refrigerated (preferred) 28 days
  Ambient  28 days
  Frozen  28 days

Clinical Information

Flecainide (Tambocor) is a Class I cardiac antiarrhythmic agent indicated for treatment of paroxysmal supraventricular dysrhythmia, paroxysmal atrial fibrillation/flutter, and life-threatening ventricular dysrhythmias. After oral administration, flecainide is almost completely absorbed and peak concentrations are attained in approximately 3 hours. The half-life averages approximately 20 hours but is widely variable (12 to 27 hours), and steady-state concentrations are typically achieved in approximately 5 days. Flecainide is eliminated from blood by hepatic metabolism, as well as renal clearance; significant changes in either organ system will cause impaired clearance. Common adverse effects include dizziness, visual disturbances, and dyspnea. Mild-to-moderate toxicity is associated with dizziness, visual disturbances, headache, nausea, fatigue, palpitations, and chest pain. Visual hallucinations and dysarthria may occur at toxic serum concentrations. Death can occur from hypotension, respiratory failure, and asystole.

Reference Values

Trough Value

0.2-1.0 mcg/mL: Therapeutic concentration

>1.0 mcg/mL: Toxic concentration

Cautions

Specimens that are obtained from gel tubes or anticoagulate collections can cause assay interference.

Day(s) Performed

Monday through Saturday

Report Available

2 to 5 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information

80181

NY State Approved

Yes