Test ID FPTHC PTH Accuratio Comprehensive Profile
Method Name
Immunoradiometric assay (IRMA)
Reporting Name
PTH Accuratio Comprehensive ProfileSpecimen Type
Plasma EDTADraw blood in a lavender-top (EDTA) tube(s). Spin down and send 2 mL EDTA plasma frozen in a plastic vial.
Note: Processing of samples should be completed within one hour of blood collection
Specimen Stability Information
Specimen Type | Temperature | Time |
---|---|---|
Plasma EDTA | Frozen |
Reference Values
COMPREHENSIVE PROFILE
TOTAL PTH Reference Range: 14.0 - 66.0 pg/mL
CAP PTH Reference Range: 5.0 - 39.0 pg/mL
CIP VALUE Reference Range: 2.5 - 29.0 pg/mL
CAP/CIP Reference Range: 1.1 - 6.9 ratio
Low ratio or PTH values are associated with Adynamic bone turnover status for an ESRD patient.
Total PTH:
Total Intact PTH is the sum of CAP PTH and N-truncated PTH fragment (likely 7-84) Total PTH = Intact PTH = CAP + CIP.
FDA approved.
CAP PTH:
CAP stands for Cyclase Activating PTH and is the same as Whole PTH which is 1-84 PTH and which raises bone turnover.
FDA approved.
CIP VALUE:
CIP stands for Cyclase Inactive PTH and the N-truncated PTH fragment, likely 7-84 PTH. CIP = Total PTH – CAP.
CAP/CIP:
CAP/CIP ratio is a calculated value and has been shown to be 93% predictive of bone turnover for the ESRD patient, however, bone biopsy is 100% predictive, therefore, clinical correlation is recommended.
Reference ranges were established using a 95% reference interval on a normal population with normal functioning kidneys. These reference ranges should not be applied to ESRD patients with impaired or non-functioning renal function.
Day(s) Performed
Varies
Report Available
5 daysPerforming Laboratory
Scantibodies Clinical LaboratoryCPT Code Information
83970\x2