Sign in →

Test ID RRBP Retinol-Binding Protein, Random, Urine

Useful For

Assessing renal tubular injury or dysfunction

 

Screening for other tubular abnormalities

 

Detecting chronic asymptomatic renal tubular dysfunction(2)

Profile Information

Test ID Reporting Name Available Separately Always Performed
CTURA Creatinine Conc Yes, (order CTU) Yes
RBPR Retinol-Binding Protein, Random, U No Yes

Method Name

CTURA: Enzymatic Colorimetric Assay

RBPR: Immunonephelometry

Reporting Name

Retinol-Binding Protein, Random, U

Specimen Type

Urine

Container/Tube: Plastic, 5-mL tube (Supply T465)

Specimen Volume: 5 mL

Collection Instructions:

1. Collect a random urine specimen.

2. No preservative.

Specimen Minimum Volume

1 mL

Specimen Stability Information

Specimen Type Temperature Time
Urine Refrigerated (preferred) 7 days
  Frozen  7 days

Clinical Information

Retinol-binding protein is a low-molecular-weight protein of 21 kDa that transports retinol (vitamin A alcohol) from the liver to peripheral tissues.(1) Retinol-binding protein is most often found bound to transthyretin, but a small, unbound fraction (<10%) passes freely through glomerular membranes and is reabsorbed by renal proximal tubules cells where it is catabolized. Due to extensive tubular reabsorption, under normal conditions very little of the filtered retinol-binding protein appears in the final excreted urine. Therefore, an increase in the urinary excretion of retinol-binding protein indicates proximal tubule injury and/or impaired proximal tubular function.(1) Measurement of retinol-binding protein in urine is, therefore, a useful aid in the monitoring and/or diagnosis of kidney disease.

 

Elevated excretion rates can indicate tubular damage associated with renal tubulointerstitial nephritis or tubular toxicity from heavy metal or nephrotoxic drug exposure. Glomerulonephropathies and renal vasculopathies also are often associated with coexisting tubular injury and so may result in elevated retinol-binding protein excretion. Measurement of urinary excretion of alpha-1-microglobulin, another low-molecular-weight protein, is an alternative to the measurement of retinol-binding protein. To date, there are no convincing studies to indicate that 1 test has better clinical utility than the other.

 

Urinary excretion of retinol-binding protein can be determined from either a 24-hour collection or from a random urine collection. The 24-hour collection is traditionally considered the gold standard. For random or spot collections, the concentration of retinol-binding protein is divided by the urinary creatinine concentration. This corrected value adjusts retinol-binding protein for variabilities in urine concentration.

Reference Values

<50 years: <130 mcg/g creatinine

≥50 years: <172 mcg/g creatinine

Cautions

Since this is a nephelometric assay, turbidity and particles (eg, cells, crystals) in the specimen can interfere with the test. Therefore, all urine specimens should be centrifuged at ambient temperature prior to assay.

Day(s) Performed

Varies; 8 a.m.-4 p.m.

Report Available

7 days

Performing Laboratory

Mayo Medical Laboratories in Rochester

Test Classification

This test uses a reagent or kit labeled by the manufacturer as Research Use Only. Its performance characteristics were determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the U.S. Food and Drug Administration.

CPT Code Information

83883

NY State Approved

Yes