Test ID SLFAT Cryptococcus Antigen Titer, Lateral Flow Assay, Serum


Specimen Required


Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 0.5 mL

Collection Instructions: Centrifuge and aliquot serum into a plastic vial


Forms

If not ordering electronically, complete, print, and send Infectious Disease Serology Test Request (T916) with the specimen.

Useful For

Monitoring Cryptococcus antigen titers in serum

 

Aiding in the diagnosis of cryptococcosis

 

This test should not be used as a test of cure or to guide treatment decisions.

Method Name

Lateral Flow Assay (LFA)

Reporting Name

Cryptococcus Ag Titer, LFA, S

Specimen Type

Serum

Specimen Minimum Volume

0.3 mL

Specimen Stability Information

Specimen Type Temperature Time
Serum Refrigerated (preferred) 14 days
  Frozen  14 days

Clinical Information

Cryptococcosis is an invasive fungal infection caused by Cryptococcus neoformans or Cryptococcus gattii. C neoformans has been isolated from several sites in nature, particularly weathered pigeon droppings. C gattii was previously only associated with tropical and subtropical regions. More recently, however, this organism has been found to be endemic in British Columbia and the Pacific Northwestern United States and is associated with several different tree species.

 

Infection is usually acquired via the pulmonary route. Patients are often unaware of any exposure history. Approximately half of the patients with symptomatic disease have a predisposing immunosuppressive condition such as AIDS, steroid therapy, lymphoma, or sarcoidosis. Symptoms may include fever, headache, dizziness, ataxia, somnolence, and cough. While the majority of C neoformans infections occur in immunocompromised patient populations, C gattii has a higher predilection for infection of healthy individuals.(1,2)

 

In addition to the lungs, cryptococcal infections frequently involve the central nervous system (CNS), particularly in patients infected with HIV. Mortality among patients with CNS cryptococcosis may approach 25% despite antibiotic therapy. Untreated CNS cryptococcosis is invariably fatal. Disseminated disease may affect any organ system and usually occurs in immunosuppressed individuals.

Reference Values

Negative 

Reference values apply to all ages.

Cautions

Cryptococcus antigen titers acquired by the lateral flow assay (LFA) may be higher than titers achieved by other Cryptococcus antigen assays. Titers acquired by different assay methods are not interchangeable.

 

Cryptococcus antigen titers should be followed using the same assay.

 

A positive result is indicative of cryptococcosis; however, all test results should be reviewed in light of other clinical findings.

 

Testing should not be performed as a screening procedure for the general populations and should only be performed when clinical evidence suggests the diagnosis of cryptococcal disease.

 

Testing hemolyzed serum samples may lead to false-negative results due to the high background color on the lateral flow assay strip.

 

Although rare, extremely high concentrations of cryptococcal antigen can result in weak test lines and in extreme instances, yield negative test results.

 

This assay has not been evaluated for cross reactivity in patients with trichosporonosis.

Day(s) Performed

Monday through Sunday

Report Available

Same day/1 to 2 days

Performing Laboratory

Mayo Clinic Laboratories in Rochester

Test Classification

This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information

87899

NY State Approved

Yes